Project Description
Application Area
Widely used in groundwater, surface water, municipal sewage, industrial sewage, environmental monitoring, and other fields water organic and inorganic phosphorus measurements.
Instrument Characteristics
Uses imported high-performance, long-life (100000 hours), high brightness light source with high stability
High precision and repeatability.
The digester adopts a 3.5-inch color touch screen with a resolution of 320×240, which is displayed in Chinese and English.
Simple and time-saving operation. The digestion colorimetry does not need to be changed.
The digester has the function of timing reminder, which can automatically shutdown after digestion, and the instrument has the function of secondary protection. When the temperature of the digester exceeds 200 ℃, the instrument will cut off the power supply by itself to ensure the safety of the digestion process.
With multi-point error free calibration, zero error measurement of set point is realized, the error caused by the nonlinearity of the temperature sensor in the common digester in the market is completely eliminated.
It can save 10 working curves and 500 historical records and data will not lose even when power is cut off.
With USB interface, data can be transferred to computer.
With printing function: you can print the measurement immediately or query and print records in the future.
Technical Indicators
【Host】 | ||||
Measurement Range: | Total phosphorus: 0.02 ~ 10mg / L | Indication Error: | ≤ ± 5% (F.S) or the largest of 0.2mg / L | |
Repeatability: | ≤3% | |||
Digestion Temperature: | 120 ° C ± 1.0 ° C | |||
Digestion Time: | 30 minutes | |||
Optical Stability: | ≤0.002A/20min | |||
Batch Volume: | 16 water samples | |||
Power Consumption: | 400W | |||
Dimensions: | Main unit 266mm × 200mm × 130mm;Digestion instrument 355mm × 260mm × 135mm | |||
【Packing List】 | ||||
Name | Quantity | Name | Quantity | |
Host 1 | 16 hole digester 1 set | |||
10 digestion colorimetric tubes | 1 set of total phosphorus reagent (100 samples can be measured) | |||
1 tube rack | 1 protective cover | |||
1 power cord | 1 operation manual | |||
1 Certificate | 1 warranty card |